RT Journal Article SR Electronic T1 Validity of the BodyGem calorimeter and prediction equations for the assessment of resting energy expenditure in overweight and obese Saudi males. JF Saudi Medical Journal JO Saudi Med J FD Prince Sultan Military Medical City SP 718 OP 724 VO 32 IS 7 A1 Almajwal, Ali M. A1 Williams, Peter G. A1 Batterham, Marijka J. YR 2011 UL http://smj.org.sa/content/32/7/718.abstract AB OBJECTIVES: To assess the accuracy of resting energy expenditure (REE) measurement in a sample of overweight and obese Saudi males, using the BodyGem device (BG) with whole room calorimetry (WRC) as a reference, and to evaluate the accuracy of predictive equations.METHODS: Thirty-eight subjects (mean ± SD, age 26.8± 3.7 years, body mass index 31.0± 4.8) were recruited during the period from 5 February 2007 to 28 March 2008. Resting energy expenditure was measured using a WRC and BG device, and also calculated using 7 prediction equations. Mean differences, bias, percent of bias (%bias), accurate estimation, underestimation and overestimation were calculated.RESULTS: Repeated measures with the BG were not significantly different (accurate prediction: 81.6%; %bias 1.1± 6.3, p>0.24) with limits of agreement ranging from +242 to -200 kcal. Resting energy expenditure measured by BG was significantly less than WRC values (accurate prediction: 47.4%; %bias: 11.0± 14.6, p = 0.0001) with unacceptably wide limits of agreement. Harris-Benedict, Schofield and World Health Organization equations were the most accurate, estimating REE within 10% of measured REE, but none seem appropriate to predict the REE of individuals.CONCLUSIONS: There was a poor agreement between the REE measured by WRC compared to BG or predictive equations. The BG assessed REE accurately in 47.4% of the subjects on an individual level.