PT - JOURNAL ARTICLE AU - Abdul-Rasheed, Omar F. AU - Farid, Yahya Y. TI - Development of a new high performance liquid chromatography method for measurement of coenzyme Q10 in healthy blood plasma DP - 2009 Sep 01 TA - Saudi Medical Journal PG - 1138--1143 VI - 30 IP - 9 4099 - http://smj.org.sa/content/30/9/1138.short 4100 - http://smj.org.sa/content/30/9/1138.full SO - Saudi Med J2009 Sep 01; 30 AB - OBJECTIVE: To improve a new high performance liquid chromatography (HPLC) method for coenzyme Q10 (CoQ10) measurement in human plasma.METHODS: We conducted this study in the Chemistry and Biochemistry Department, College of Medicine, Al-Nahrain University, Baghdad, Iraq from September 2004 to July 2005. Sixty healthy volunteers aged 28.51±7.30 years were participated in this study. All participants were non-smokers and under no any treatment for the last 3 days and fast for 12 hours before analysis. Coenzyme Q10 levels were measured by improved isocratic reversed phase HPLC technique after one-step ice-cold 1-propanol extraction step.RESULTS: The ultra violet (UV) detection, developed HPLC method gives a good linearity range between 0.1-4.0 mg/L. The coefficients of variations (CVs%) of within-day precision for CoQ10 in human plasma were 0.2-3.90%. Analytical recoveries were 95.5-101.30%, whereas the average CV% of CoQ10 reproducibility was 0.959%.CONCLUSION: A new improved rapid, precise, and sensitive HPLC method for the quantitation of CoQ10 is suitable for analysis of blood plasma samples.