PT - JOURNAL ARTICLE AU - Alrehily, Faisal A. AU - Alqahtani, Khalid S. AU - Aljohani, Marwan H. AU - Alharbi, Nawaf S. AU - Alharbi, Reyan M. AU - Abdulaal, Osamah M. AU - Alshamrani, Abdullah F. AU - Alsaedi, Abdulrahman S. AU - Al-Murshedi, Sadeq H. AU - Alhazmi, Fahad H. TI - Establishing local diagnostic reference levels for computed tomography examinations using size-specific dose estimates AID - 10.15537/smj.2023.44.8.20230230 DP - 2023 Aug 01 TA - Saudi Medical Journal PG - 761--766 VI - 44 IP - 8 4099 - http://smj.org.sa/content/44/8/761.short 4100 - http://smj.org.sa/content/44/8/761.full SO - Saudi Med J2023 Aug 01; 44 AB - Objectives: To establish local DRL (LDRL) for computed tomography (CT) examinations based on size-specific dose estimates (SSDEs), which consider patient size. The concept of diagnostic reference level (DRL) was introduced to limit patient exposure to unnecessary radiation. However, traditional DRL values do not consider patient size.Methods: Following institutional committee approval, data were collected from CT examinations of adult patients at Madinah General Hospital, Al Madinah Al Munawwarah, Saudi Arabia from January to March 2023. The SSDE was calculated for each patient using the effective diameter (Deff).Results: The LDRLs of the brain, cervical spine, chest, thoracic spine and kidneys, ureters, and bladder (KUB) examinations were 118 mGy, 12 mGy, 8 mGy, 17 mGy, and 7 mGy, respectively. A strong correlation was observed between SSDEs and the volume computed tomography dose index (CTDIvol) for all examinations except chest scans (p<0.05). Size-specific dose estimates were higher than the CTDIvol, with a greater difference for patients with smaller Deff (p<0.05).Conclusion: The established LDRL was within the international DRL. The use of SSDE has the potential to provide more accurate and relevant data for radiation safety practices; however, widespread adoption of SSDE in new CT scanners is necessary for promoting consistency and standardization methodologies.