Precision of the PRECICE® Internal Bone Lengthening Nail

  • Symposium: 2013 Limb Lengthening and Reconstruction Society
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Previous designs of internal bone lengthening devices have been fraught with imprecise distraction, resulting in nerve injuries, joint contractures, nonunions, and other complications. Recently, a magnet-operated PRECICE® nail (Ellipse Technologies, Inc, Irvine, CA, USA) was approved by the FDA; however, its clinical efficacy is unknown.

Questions/purposes

We evaluated this nail in terms of (1) accuracy and precision of distraction, (2) effects on bone alignment, (3) effects on adjacent-joint ROM, and (4) frequency of implant-related and non-implant-related complications.

Methods

We reviewed medical and radiographic records of 24 patients who underwent femoral and/or tibial lengthening procedures using the PRECICE® nail from August 2012 to July 2013 for conditions of varied etiology, the most common being congenital limb length discrepancy, posttraumatic growth arrest, and fracture malunion. This group represented 29% of patients (24 of 82) who underwent a limb lengthening procedure for a similar diagnosis during the review period. At each postoperative visit, the accuracy and precision of distraction, bone alignment, joint ROM, and any complications were recorded by the senior surgeon (SRR). Accuracy reflected how close the measured lengthening was to the prescribed distraction at each postoperative visit, while precision reflected how close the repeated measurements were to each other over the course of total lengthening period. No patients were lost to followup. Minimum followup from surgery was 3 weeks (mean, 14 weeks; range, 3–29 weeks).

Results

Mean total lengthening was 35 mm (range, 14–65 mm), with an accuracy of 96% and precision of 86%. All patients achieved target lengthening with minimal unintentional effects on bone alignment. The knee and ankle ROM were minimally affected. Of the complications requiring return to the operating room for an additional surgical procedure, there was one (4%) implant failure caused by a nonfunctional distraction mechanism and six (24%) non-implant-related complications, including premature consolidation in one patient (4%), delayed bone healing in two (8%), delayed equinus contracture in two (8%), and toe clawing in one (4%).

Conclusions

We conclude that this internal lengthening nail is a valid option to achieve accurate and precise limb lengthening to treat a variety of conditions with limb shortening or length discrepancy. Randomized, larger-sample, long-term studies are required to further confirm clinical efficacy of these devices, monitor for any late failures and complications, and compare with other internal lengthening devices with different mechanisms of operation.

Level of Evidence

Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3A–B
Fig. 4
Fig. 5A–C
Fig. 6

Similar content being viewed by others

References

  1. Baumann F, Harms J. [The extension nail: a new method for lengthening of the femur and tibia (author’s transl)] [in German]. Arch Orthop Unfallchir. 1977;90:139–146.

    Article  CAS  PubMed  Google Scholar 

  2. Baumgart R, Betz A, Schweiberer L. A fully implantable motorized intramedullary nail for limb lengthening and bone transport. Clin Orthop Relat Res. 1997;343:135–143.

    Article  PubMed  Google Scholar 

  3. Baumgart R, Thaller P, Hinterwimmer S, Krammer M, Hierl T, Mutschler W. A fully implantable, programmable distraction nail (Fitbone)—new perspectives for corrective and reconstructive limb surgery. In: Kempf I, Leung KS, Alt V, Haarman H, Taglang G, Schnettler R, Seidel H. eds. Practice of Intramedullary Locked Nail: New Developments in Techniques and Applications. New York, NY: Springer; 2006:189–198.

    Chapter  Google Scholar 

  4. Bliskunov AI. [Implantable devices for lengthening the femur without external drive mechanisms] [in Russian]. Med Tekh. 1984;2:44–49.

    PubMed  Google Scholar 

  5. Bost FC, Larsen LJ. Experiences with lengthening of the femur over n intramedullary rod. J Bone Joint Surg Am. 1956;38:567–584.

    PubMed  Google Scholar 

  6. Burghardt RD, Herzenberg JE, Specht SC, Paley D. Mechanical failure of the Intramedullary Skeletal Kinetic Distractor in limb lengthening. J Bone Joint Surg Br. 2011;93:639–643.

    Article  CAS  PubMed  Google Scholar 

  7. Burghardt RD, Paley D, Specht SC, Herzenberg JE. The effect on mechanical axis deviation of femoral lengthening with an intramedullary telescopic nail. J Bone Joint Surg Br. 2012;94:1241–1245.

    Article  CAS  PubMed  Google Scholar 

  8. Cheung KM, Cheung JP, Samartzis D, Mak KC, Wong YW, Cheung WY, Akbarnia BA, Luk KD. Magnetically controlled growing rods for severe spinal curvature in young children: a prospective case series. Lancet. 2012;379:1967–1974.

    Article  PubMed  Google Scholar 

  9. Cole JD, Justin D, Kasparis T, DeVlught D, Knobloch C. The Intramedullary Skeletal Kinetic Distractor (ISKD): first clinical results of a new intramedullary nail for lengthening of the femur and tibia. Injury. 2001;32:SD129–SD139.

    Article  PubMed  Google Scholar 

  10. Dahl MT, Gulli B, Berg T. Complications of limb lengthening: a learning curve. Clin Orthop Relat Res. 1994;301:10–18.

    PubMed  Google Scholar 

  11. Dincyurek H, Kocaoglu M, Eralp IL, Bilen FE, Dikmen G, Eren I. Functional results of lower extremity lengthening by motorized intramedullary nails. Acta Orthop Traumatol Turc. 2012;46:42–49.

    Article  PubMed  Google Scholar 

  12. Ellipse Technologies Inc. PRECICE Nail System: Ellipse Technologies. Available at: http://www.ellipse-tech.com/?q=ip. Accessed December 15, 2012.

  13. Faber FW, Keessen W, van Roermund PM. Complications of leg lengthening. 46 procedures in 28 patients. Acta Orthop Scand. 1991;62:327–332.

    Article  CAS  PubMed  Google Scholar 

  14. Garcia-Cimbrelo E, Curto de la Mano A, Garcia-Rey E, Cordero J, Marti-Ciruelos R. The intramedullary elongation nail for femoral lengthening. J Bone Joint Surg Br. 2002;84:971–977.

    Article  CAS  PubMed  Google Scholar 

  15. Guichet JM, Casar RS. Mechanical characterization of a totally intramedullary gradual elongation nail. Clin Orthop Relat Res. 1997;337:281–290.

    Article  PubMed  Google Scholar 

  16. Guichet JM, Deromedis B, Donnan LT, Peretti G, Lascombes P, Bado F. Gradual femoral lengthening with the Albizzia intramedullary nail. J Bone Joint Surg Am. 2003;85:838–848.

    PubMed  Google Scholar 

  17. Guichet JM, Grammont PM, Trouilloud P. [A nail for progressive lengthening: an animal experiment with a 2-year follow-up] [in French]. Chirurgie. 1992;118:405–410.

    CAS  PubMed  Google Scholar 

  18. Hankemeier S, Gosling T, Pape HC, Wiebking U, Krettek C. Limb lengthening with the Intramedullary Skeletal Kinetic Distractor (ISKD). Oper Orthop Traumatol. 2005;17:79–101.

    Article  PubMed  Google Scholar 

  19. Harbacheuski R, Fragomen AT, Rozbruch SR. Does lengthening and then plating (LAP) shorten duration of external fixation? Clin Orthop Relat Res. 2012;470:1771–1781.

    Article  PubMed Central  PubMed  Google Scholar 

  20. Herzenberg JE, Scheufele LL, Paley D, Bechtel R, Tepper S. Knee range of motion in isolated femoral lengthening. Clin Orthop Relat Res. 1994;301:49–54.

    PubMed  Google Scholar 

  21. Ilizarov GA. The tension-stress effect on the genesis and growth of tissues. Part II. The influence of the rate and frequency of distraction. Clin Orthop Relat Res. 1989;239:263–285.

    PubMed  Google Scholar 

  22. Kenawey M, Krettek C, Liodakis E, Meller R, Hankemeier S. Insufficient bone regenerate after intramedullary femoral lengthening: risk factors and classification system. Clin Orthop Relat Res. 2011;469:264–273.

    Article  PubMed Central  PubMed  Google Scholar 

  23. Kenawey M, Krettek C, Liodakis E, Wiebking U, Hankemeier S. Leg lengthening using Intramedullary Skeletal Kinetic Distractor: results of 57 consecutive applications. Injury. 2011;42:150–155.

    Article  PubMed  Google Scholar 

  24. Konofaos P, Kashyap A, Neel MD, Ver Halen JP. A novel device for long bone osteodistraction: description of device and case series. Plast Reconstr Surg. 2012;130 418e–422e.

    Article  CAS  PubMed  Google Scholar 

  25. Krettek C, Miclau T, Schandelmaier P, Stephan C, Mohlmann U, Tscherne H. The mechanical effect of blocking screws (“Poller screws”) in stabilizing tibia fractures with short proximal or distal fragments after insertion of small-diameter intramedullary nails. J Orthop Trauma. 1999;13:550–553.

    Article  CAS  PubMed  Google Scholar 

  26. Krieg AH, Lenze U, Speth BM, Hasler CC. Intramedullary leg lengthening with a motorized nail. Acta Orthop. 2011;82:344–350.

    Article  PubMed Central  PubMed  Google Scholar 

  27. Krieg AH, Speth BM, Foster BK. Leg lengthening with a motorized nail in adolescents: an alternative to external fixators? Clin Orthop Relat Res. 2008;466:189–197.

    Article  PubMed Central  PubMed  Google Scholar 

  28. Leidinger B, Winkelmann W, Roedl R. [Limb lengthening with a fully implantable mechanical distraction intramedullary nail] [in German]. Z Orthop Ihre Grenzgeb. 2006;144:419–426.

    Article  CAS  PubMed  Google Scholar 

  29. Mahboubian S, Seah M, Fragomen AT, Rozbruch SR. Femoral lengthening with lengthening over a nail has fewer complications than intramedullary skeletal kinetic distraction. Clin Orthop Relat Res. 2012;470:1221–1231.

    Article  PubMed Central  PubMed  Google Scholar 

  30. Mazeau P, Assi C, Louahem D, L’Kaissi M, Delpont M, Cottalorda J. Complications of Albizzia femoral lengthening nail: an analysis of 36 cases. J Pediatr Orthop B. 2012;21:394–399.

    Article  PubMed  Google Scholar 

  31. Noonan KJ, Leyes M, Forriol F, Canadell J. Distraction osteogenesis of the lower extremity with use of monolateral external fixation: a study of two hundred and sixty-one femora and tibiae. J Bone Joint Surg Am. 1998;80:793–806.

    CAS  PubMed  Google Scholar 

  32. Paley D. Problems, obstacles, and complications of limb lengthening by the Ilizarov technique. Clin Orthop Relat Res. 1990;250:81–104.

    PubMed  Google Scholar 

  33. Paley D, Herzenberg JE, Paremain G, Bhave A. Femoral lengthening over an intramedullary nail: a matched-case comparison with Ilizarov femoral lengthening. J Bone Joint Surg Am. 1997;79:1464–1480.

    CAS  PubMed  Google Scholar 

  34. Paley D, Tetsworth K. Percutaneous osteotomies: osteotome and Gigli saw techniques. Orthop Clin North Am. 1991;22:613–624.

    CAS  PubMed  Google Scholar 

  35. Papanna MC, Monga P, Al-Hadithy N, Wilkes RA. Promises and difficulties with the use of femoral intra-medullary lengthening nails to treat limb length discrepancies. Acta Orthop Belg. 2011;77:788–794.

    PubMed  Google Scholar 

  36. Rozbruch SR, Kleinman D, Fragomen AT, Ilizarov S. Limb lengthening and then insertion of an intramedullary nail: a case-matched comparison. Clin Orthop Relat Res. 2008;466:2923–2932.

    Article  PubMed Central  PubMed  Google Scholar 

  37. Sabharwal S, Rozbruch SR. What’s new in limb lengthening and deformity correction? J Bone Joint Surg Am. 2011;93:2323–2332.

    PubMed  Google Scholar 

  38. Schiedel FM, Pip S, Wacker S, Popping J, Tretow H, Leidinger B, Rodl R. Intramedullary limb lengthening with the Intramedullary Skeletal Kinetic Distractor in the lower limb. J Bone Joint Surg Br. 2011;93:788–792.

    Article  CAS  PubMed  Google Scholar 

  39. Simpson AH, Kenwright J. Fracture after distraction osteogenesis. J Bone Joint Surg Br. 2000;82:659–665.

    Article  CAS  PubMed  Google Scholar 

  40. Simpson AH, Shalaby H, Keenan G. Femoral lengthening with the Intramedullary Skeletal Kinetic Distractor. J Bone Joint Surg Br. 2009;91:955–961.

    Article  CAS  PubMed  Google Scholar 

  41. Tjernstrom B, Olerud S, Rehnberg L. Limb lengthening by callus distraction: complications in 53 cases operated 1980–1991. Acta Orthop Scand. 1994;65:447–455.

    Article  CAS  PubMed  Google Scholar 

  42. Wang K, Edwards E. Intramedullary Skeletal Kinetic Distractor in the treatment of leg length discrepancy—a review of 16 cases and analysis of complications. J Orthop Trauma. 2012;26 e138–e144.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We thank the Ellipse Technologies sales representatives and engineers for technical guidance.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to S. Robert Rozbruch MD.

Additional information

Each author certifies that he or she, or a member of his or her immediate family, has no funding or commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc) that might pose a conflict of interest in connection with the submitted article.

All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research ® editors and board members are on file with the publication and can be viewed on request.

Clinical Orthopaedics and Related Research ® neither advocates nor endorses the use of any treatment, drug, or device. Readers are encouraged to always seek additional information, including FDA approval status, of any drug or device before clinical use.

Each author certifies that his or her institution approved the human protocol for this investigation, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

About this article

Cite this article

Kirane, Y.M., Fragomen, A.T. & Rozbruch, S.R. Precision of the PRECICE® Internal Bone Lengthening Nail. Clin Orthop Relat Res 472, 3869–3878 (2014). https://doi.org/10.1007/s11999-014-3575-0

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-014-3575-0

Keywords

Navigation